BEEM 100 mA – SINGLE TANK
100 KVP DIGITAL DIAGNOSTIC X-RAY UNIT AS PER ‘IS’ 7620-1986
FEATURES
| • |
ON & OFF Push button power switches |
| • |
Manual Voltage compensation in coarse and fine steps up to 80 volts |
| • |
15 position rotary switch for selection of radiography & fluoroscopy KVP from 40 to 100 KVP |
| • |
Rotary Selection Switch for different mA: fluoro 30 mA, 60 mA and 100 mA |
| • |
24 position switch for exposure time from 0.02 to 5 sec |
| • |
Toggle switch for Bucky on / 0ff |
| • |
Stepless mA selector for setting of Fluoro mA values from 0 to 5mA |
| • |
Digital display for radiography film exposure parameters |
| • |
Green light indicator for exposure |
| • |
Yellow light indicator for overload |
| • |
Red light indicator for power supply |
| • |
Actual radiography analog mA meter from 0 to 100 mA |
| • |
Actual fluoro analogue mA meter from 0 to 10 mA |
| • |
Digital display of mA, KVP and mAs |
| • |
Automatic Calculation of mAs value as per mA setting |
| TECHNICAL DATA |
BEEM 100 |
Power |
220 220 V ± 10 % |
| Frequency |
50 Hz ± 1 Hz |
| Power consumption |
7.5 KVA |
| Inner resistance |
1.5 Ohms |
| Radiographic KVP |
40 to 100 KV in 15 steps |
| Fluoroscopic KVP |
40 to 90 KV in 15 steps |
| Radiography mA |
30, 60, 100 mA |
| Fluoroscopy mA |
0 to 5 mA |
| mAs range |
0.6 to 240 mAs in 24 steps |
| Exposure time |
20ms – 5s |
| X-ray tube |
Stationary Anode |
| Focal spot |
Single focus 2.6 x 2.6 mm |
| Safety Standard |
IS: 7620 |
5-D Table
| • |
Hand operated five-position table including trendelenberg |
| • |
Counter balanced bucky radiography in all positions of table |
| • |
Sturdy construction |
| • |
Simple operation |
| • |
Counter balanced full travel motorized bucky with high ratio grid |
| • |
Stainless steel cassette tray to take cassettes up to 14”x17” size (Dimension: 720 mmW / 850 mmH / 1830 mmL) |
COLUMN STAND
| • |
Floor to ceiling balanced column stand for x-ray tube for effortless movement in all directions with necessary locks. |
| • |
Column stand adjustable to a ceiling height of 8.5 – 11 feet and 360° rotatable |
| • |
Lateral exposure facility |
| • |
Easy and firm foot lock for floor trowel. |
L.B.D
| • |
Double slot manual collimator with bright halogen light |
OPTIONAL
| • |
Screening and spot film facility |
| • |
Horizontal table (optional attachment) |
APPLICATION
Routine Radiography, Fluoroscopy, Bucky Radiography and Chest Radiography